Introduction
Filter mesh test stand is based on relevant standards and specification to test quality control factors: pipe leakage, dust generation, flow test, resistance test, the change of different filters, flow control.
Structure
· Power section
· Aerosol generator section
The aerosol generator can consecutive produce stable DEHS particles (particle size range 0.3-3μm) .
· Particle counter section
The sampling flow of the particle counter is 2.83L/min. The particle counter is specially suitable for particle size 0.3μm-10μm.
· Fixture and structure section
The pipeline is made of 305 stainless steel, anticorrosion and antirust, with consideration of convenient pipeline replacement ( when testing different size of filters).
· Flow monitoring section: Using V cone flowmeter for flow detection.
· Differential pressure monitoring section:
Using imported DWYER electronic differential pressure transducer for pressure measurement.
· Flow control section
Electric control valve to automatically regulate the flow
· Data processing:
Including PLC, software and electric control
Standards
ISO 11155 Road vehicles - Air filters for passenger compartments - Part 2: Test for gaseous filtration
QC/T 795-1:2007 Road vehicles - Air filters for passenger compartments - Part 1: Dust filtration test.
Technical Parameters
Item |
Parameters |
Basic Parameter |
Power supply |
Three phase AC, 380V±5% 50Hz±2Hz ,5.5kw |
Condensed air |
4-8kg/cm2 |
Dust source |
DEHS |
Anti-electric shock Rate |
Grade I |
Environment condition |
Temperature |
15-25 ℃ |
Relative humidity |
≤75%RH |
Air pressure |
68kPa-106kPa |
Flow |
Test range |
150-600m³/h |
Test accuracy |
±5% |
Test stability |
±3% |
Resistance |
Test range |
0-250pa |
Test accuracy |
0-100pa,±5pa;>100pa ,±5% |
Aerosol |
Uniformity |
≤±15% |
Particle counter calibration |
Calibrated in according to JIF 1190-2008 (particle counter calibration specification ) |
Repeatability of counting efficiency |
Efficiency lower than 40%, ≤±15%;
Efficiency higher than 40%, ≤±5% |