Textile Thermal Resistance Tester

Standards:
ISO 11092
ASTM D1518
ASTM F1868
GB/T 11048
Scope of Application:
Research and Development: To evaluate the thermal performance of new fibers, yarns, fabric constructions, and multilayer systems.
Quality Control: To ensure manufactured textile materials meet specified thermal resistance requirements consistently.
Product Performance Evaluation: To compare the thermal insulation of different textile products or assess the impact of aging, washing, or other factors on thermal performance.
Compliance Testing: To provide data required for product labeling and certification according to relevant thermal comfort or protective clothing standards.
Product Advantages:
The Textile Thermal Resistance Tester offers several advantages for evaluating the thermal insulation of textiles:
Measures Fundamental Property: Directly quantifies thermal resistance, a key indicator of a textile's insulating capability and contribution to thermal comfort.
Simulates Heat Transfer: Replicates the heat exchange processes (conduction, convection, radiation) that occur in real-world clothing use.
Provides Quantitative Data: Yields objective and reproducible numerical values for thermal resistance.
Product Features:
A constant-temperature heat source (hot plate) with precise temperature control to simulate skin temperature.
A guarding system around the hot plate to minimize lateral heat loss and ensure one-dimensional heat flow through the test specimen.
A cooling plate or controlled environment chamber to simulate the external temperature.
A mechanism for clamping the textile sample flatly between the heat source and the cooling plate or within the controlled environment.
High-precision temperature sensors embedded in the hot plate, cooling plate, and/or environmental chamber to monitor temperatures accurately.
Technical Parameters:
1. thermal resistance (Rct) range :(0.002-2.000)m2.K/W resolution 0.0001m2.K/W
2. wet resistance (Ret) range :(1.0-1000) m2.pa /W resolution m2.pa /W
3. the test bench temperature range :(20 ~ 50)ºC adjustable; Plus or minus 0.1 ºC
4. temperature measurement accuracy: ±0.1ºC resolution: 0.01ºC
5. humidity measurement accuracy: ±3% resolution: 0.1%
6. sample thickness range :(0 ~ 70)mm
7. the airflow speed in the air duct :(1±0.05) m/s 0.1m/s ~ 2m /s(adjustable)
8. External size of thermal interring: 512mm×512mm
9. Width of thermal protection ring: 127mm
10. test plate size: 250mm x 250mm
11. sample size: 515mm×515mm
12. climate chamber size: 830mm× 780mm×630mm
13. power supply: AC380V±10% 50Hz 3.5KW
14. dimensions: host (1100×1220×1820)mm, computer (690×690×800)mm
15. weight: about 280kg
Leave Message Get Price