The characteristic of Anti Synthetic blood penetration tester
 
 
	The Synthetic Blood Penetration – Splash Resistance test is one of two synthetic blood resistance tests offered by Nelson Labs to determine a product’s ability to act as a barrier to blood-borne pathogens. The Splash Resistance test method challenges medical face masks with a fixed volume of synthetic blood directed at high velocity at the center of the mask.
	This test is required by ASTM F2100 and is tested in compliance with ASTM F1862, ASTM F2100, EN 14683, and ISO 26609. Nelson Labs’ highly experienced staff provides a one-stop shop for all ASTM F2100 and EN 14683 testing.
	 
	Applicable Standards
	ASTM F1862
	ISO 22609
	ASTM F2100
	 
	Sample Specifications
	SBP210 : 32 masks are required for each pressure
	 
	Study Outline
	2 mL of synthetic blood is sprayed through a small cannula onto the surface of the face mask. At the conclusion of the test, the back side of the medical face mask is observed for synthetic blood penetration. Face masks can be evaluated at three different velocities corresponding to a human blood pressure of 80, 120, and 160 mm Hg. If the mask passes at the highest pressure, there is no need to test the mask at a lower pressure. A single sampling plan providing Acceptable Quality Level (AQL) of 4.0% would require that 32 masks be tested and that 29 of the 32 masks pass the test.
	The surface tension range for blood and body fluids is approximately 42 – 60 dynes/cm. To simulate the wetting characteristics, the synthetic blood is adjusted to the lower end of this range (40 ± 5 dynes/cm).
	
	 
            2020-07-10 10:54