The Air Filter Performance Test Equipment is suitable for dry and wet (oil bath), pilot air filter assembly.
Technical Parameters:
Air flow range | 15-3000m3/h (Under standard atmospheric conditions) |
Air flow measurement accuracy | ±0.5% |
In front of nozzle cavity stagnation pressure measurement range | 0~110kPa |
The accuracy of the absolute pressure transmitter | ±0.25% |
In front of nozzle cavity stagnation temperature measurement range | 0~100°C |
Accuracy of temperature sensors | A grade |
Air flow measurement method: with a critical flow venturi nozzle measuring air flow rate of air flow measurement accuracy and stability of the highly critical flow venturi nozzle. System through the flow ratio of multiple relationship (1,2,4,8, ...), a number of automatic composition and pressure of the sonic nozzle automatic pressure regulating valve to adjust the air flow setting and continuously adjustable.
Powder Feeder:
The amount of powder | 1~100g/min |
Power bucket one-time Max. load amount of powder | 1000g |
Accuracy | every 5 minutes plus ash content does not change more than 5% of the average ash content |
Air intake resistance measurement:
Measuring range: 0~40kPa
Accuracy: ±0.1%
The intake resistance automatically recorded by the computer the data sampled by the display, and can draw the intake resistance curve.