Salt spray tester common causes of failure and treatment methods:
Fault one: the experiment can not rise to the set temperature
Cause analysis:
1. The temperature of the laboratory temperature controller is set too low
2. Laboratory safety protection switch is set too low
3. Electromagnetic relay failure
Solution:
1 Set the temperature controller to the desired temperature
2 Set the safety switch to the desired temperature
3 Replace the solenoid relay
Fault 2: The temperature of the saturation tank cannot rise to the set temperature
Analysis of the cause:
1Saturated tank temperature controller is too low
2Saturated hopper safety switch set too low
3Failure of the electromagnetic relay
Solution:
1Set the temperature controller to the desired temperature
2Set the safety switch to the required temperature
3Replace the solenoid relay
Fault 3: Insufficient spraying
Analysis of the cause:
1Spray regulator position too low
2Clogged preheating tank glass filter
3Pressure setting too low
Remedy:
1Open the spray regulator
2Clean the glass filter
3Regulate the regulator to 1kg/cm2, there is a regulator on the air compressor
Fault 4:When unable to spray
Cause Analysis:
1Air compressor is not running
2Air compressor outlet main switch is not open
3 Solenoid valve failure
4 pressure gauge failure or low pressure
5Nozzle blockage
Treatment method:
1、Open the air compressor button
2、Open the main switch of air compressor
3、Replace the solenoid valve
4、Replace the pressure gauge and check the air source of the compressor
5、Remove the nozzle for cleaning (please disassemble carefully)
Other situations:
1、When the low water level warning light is on, it means that the water level is too low, check whether the water source is fed with water.
2、The spray is normal, but the air compressor is not running, the air compressor itself has self-protection function, use as usual.
3、Forbidden to operate after power on
4、When the water level of the heating tank is too low, it will cut off the running power, just increase the water level of the heating tank to the normal state
5、When the temperature controller shows EEE:
a, temperature controller failure (replacement parts))
b、Temperature sensor failure (replacement parts)
6、When the spraying volume is too large:
a、Spray regulator position is too high
b、The glass nozzle is used for a long time and the diameter of the nozzle is too large
7、Spray regulator down :
a、Replacement of new glass nozzle
b, adjust the pressure to 2kg/cm2